SPECIFICATIONS
|
25-1.
Meter Dimensions
|
SIGNAL
INPUT
Isolation to Earth Ground
- 354 V peak per
IEC spacing
- NMR- 60 dB
- CMR- 120 dB
Protection
- 240 V rms max
for voltage input ranges
Display
LED 14-segment, 13.8 mm (0.54")
Symbols: 8. 8. 8. 8.
ANALOG TO DIGITAL
Technique: Dual slope
Internal resolution: 15 bits
Read Rate: 3/sec Polarity Automatic
Polarity: Automatic
DIN PlatinumTemperature range:
-200░ to 850░C (-328 to 1562░F)
Alpha = 0.00385 (DIN 43760)
ACCURACY
AT 25░C:
▒0.5░C
Temperature
Stability: ▒0.04░C/░C
Lead Resistance for Specified Accuracy
2 Wire: Up to 100 milliohm/lead
3 Wire: Up to 10 ohms/lead balanced
4 Wire: Up to 20 ohms/total unbalanced
Step
Response: 1 sec
Warm Up to Rated Accuracy: 30 min
ANALOG
OUTPUT (if applicable)
Signal Type: Current or voltage
Signal Level
Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function
May be assigned to a display range or proportional control output with
setpoint #1 when used as a control output.
Linearity: 0.2%
Load Regulation: 1.1%
Line Regulation: 0.02%/V ac
Step Response Time
2 - 3 seconds to 99% of the final value
INPUT
POWER INFORMATION
Voltage
ac - 115/230 Vrms ▒10%
dc - 9.5 at 950mA dc to 32 V dc at 250mA dc
Frequency: 50-60 Hz
Power: 6 watts max.
ENVIRONMENT
Operating Temperature:
0 to 50░C (115/230 V rms ▒15%)
0 to 60░C (115/230 V rms ▒10%)
Storage Temperature: -40░ to 85░C
Relative Humidity: 90% at 40░C (non-condensing)
MECHANICAL
Panel cutout: 1/8 DIN 3.62" x 1.78" (45 x 92mm)
Weight: 1.27 lb (575 g)
Case material: Polycarbonate, 94 V-O UL rated
ALARM
OUTPUTS (if applicable)
2 Form "C" on/off relays. Configurable for latched and unlatched by software.
Max current: 6 AMPS
Max voltage: 250 Vac or 30 Vdc
|